cic.analysisRun.computeCIC
Calculate charge-injection and voltage-transient metrics.
Syntax
A = cic.analysisRun.computeCIC(item)
A = cic.analysisRun.computeCIC(item, opts)Description
Analyzes one biphasic chronopotentiometry transient. The function detects the cathodic and anodic pulse windows, samples the polarization voltage after each phase, integrates injected charge, normalizes charge by electrode area, and checks the measured voltages against the selected cathodic and anodic limits. It performs no plotting or file export.
Inputs
item- Scalar DTA item structure. item.curve, or the main curve selected from item.tables, must provide T in seconds, Vf in volts, and Im in amperes. Pt is optional; sample numbers are generated when it is absent. item.meta may contain pulse metadata and area_cm2.
opts- Optional scalar structure. See Options.
Options
delay_s- Time in seconds from the end of each pulse phase to the Emc or Ema sample. The voltage is linearly interpolated at cath_end+delay_s and anod_end+delay_s. Default: 10e-6 seconds.
cathLimit- Cathodic water-window limit in volts. Emc passes when Emc >= cathLimit. Default: -0.6 V.
anodLimit- Anodic water-window limit in volts. Ema passes when Ema <= anodLimit. Default: 0.8 V.
areaOverride- Positive electrode area in square centimetres, supplied as a numeric scalar or numeric text. A valid value takes precedence over area_cm2 and item.meta.area_cm2. Default: "".
area_cm2- Positive numeric electrode area in square centimetres used when areaOverride is empty or invalid. If neither option supplies an area, item.meta.area_cm2 is used. Without a valid area, charge is still calculated but CIC density fields are NaN. Default: NaN.
pulseMode- Pulse-detection policy. Allowed display values are "Metadata first, then auto", "Metadata only", and "Auto from Im only". The equivalent values "metadata_first", "metadata_only", and "current_only" are also accepted by
labkit.dta.detectPulses. Default: "Metadata first, then auto". usedMeasuredCurrent- Logical scalar. When true, Qc and Qa are the absolute trapezoidal integrals of measured Im over the detected phase windows. When false, charge is the absolute nominal phase current multiplied by phase duration. Default: true.
Calculations
Samples with NaN T, Vf, or Im are removed together. At least five remaining samples are required. Emc is Vf at cath_end+delay_s and Ema is Vf at anod_end+delay_s. The cathodic baseline Eipp uses the first finite value in this order: pre-pulse median, interpulse median, post-pulse median, zero. The anodic baseline Eipp_gap prefers the interpulse median, followed by the pre-pulse median, post-pulse median, and Eipp.
Qc_C and Qa_C are stored as positive charge magnitudes in coulombs; Qt_C = Qc_C + Qa_C. For a valid area, CIC density is 1000*Q/area_cm2 in mC/cm^2. CICc_mCcm2, CICa_mCcm2, and CICt_mCcm2 refer to the cathodic, anodic, and summed biphasic charge respectively.
Outputs
A- Scalar result structure. A.ok is true only when curve selection, pulse detection, delayed voltage sampling, and both charge integrations succeed. A.message is "OK" on success and explains the first failed stage otherwise.
Output Fields
ok- Logical success flag.
message- "OK" or a description of the first failed stage.
delay_s- Effective delayed-voltage sampling interval in seconds.
cathLimit- Effective cathodic voltage limit in volts.
anodLimit- Effective anodic voltage limit in volts.
area_cm2- Selected positive electrode area in cm^2, or NaN when missing.
usedMeasuredCurrent- Effective charge-integration policy.
logOnFailure- True for failures that the app should copy into its visible diagnostic log; false for ordinary incomplete-input results.
Emc- Delayed cathodic polarization voltage in volts.
Ema- Delayed anodic polarization voltage in volts.
t_emc- Requested Emc sample time in seconds.
t_ema- Requested Ema sample time in seconds.
emc_idx- Index of the recorded sample nearest t_emc.
ema_idx- Index of the recorded sample nearest t_ema.
Epre- Median pre-pulse voltage in volts.
Ebetween- Median interpulse voltage in volts.
Epost- Median post-pulse voltage in volts.
Eipp- Baseline selected for the cathodic phase, in volts.
Eipp_gap- Baseline selected for the anodic phase, in volts.
baselineCathSource- Text naming the cathodic baseline source.
baselineAnodSource- Text naming the anodic baseline source.
baselineCathWindow- Two-element time window used for Eipp, in seconds.
baselineAnodWindow- Two-element time window used for Eipp_gap, in seconds.
Vc_on- Voltage sampled delay_s after cathodic phase onset, in volts.
Va_on- Voltage sampled delay_s after anodic phase onset, in volts.
t_conset- Requested cathodic onset sample time in seconds.
t_aonset- Requested anodic onset sample time in seconds.
Va_cath_mag- Absolute Eipp-to-Vc_on voltage change, in volts.
Va_anod_mag- Absolute Eipp_gap-to-Va_on voltage change, in volts.
tc_s- Cathodic phase duration in seconds.
ta_s- Anodic phase duration in seconds.
tip_s- Interpulse gap duration in seconds.
Ic_est_A- Median measured cathodic current in amperes.
Ia_est_A- Median measured anodic current in amperes.
cathMask- Logical vector selecting cathodic phase samples.
anodMask- Logical vector selecting anodic phase samples.
Qc_C- Cathodic charge magnitude in coulombs.
Qa_C- Anodic charge magnitude in coulombs.
Qt_C- Qc_C + Qa_C in coulombs.
CICc_mCcm2- Cathodic charge density in mC/cm^2.
CICa_mCcm2- Anodic charge density in mC/cm^2.
CICt_mCcm2- Total biphasic charge density in mC/cm^2.
cathOK- True when Emc is not below cathLimit.
anodOK- True when Ema is not above anodLimit.
safe- True when both cathOK and anodOK are true.
limitSide- "safe", "cathodic exceeded", "anodic exceeded", or "both exceeded".
pulse- Pulse geometry returned by
labkit.dta.detectPulses. detectMode- Pulse-detection method that produced pulse.
detectMsg- Pulse-detection status message.
t- Filtered time vector in seconds.
Vf- Filtered voltage vector in volts.
Im- Filtered current vector in amperes.
pt- Filtered source point numbers or generated zero-based sample numbers.
sample_dt- Median difference between consecutive filtered times, in seconds. sample_dt_report contains the same value for exports.
ampEstimate_A- Largest absolute filtered current in amperes.
Failure Behavior
Missing curves, too few valid samples, failed pulse detection, an out-of-range delay, or a pulse window with fewer than two samples returns A.ok=false. These data failures normally return a message instead of throwing. Invalid MATLAB types or malformed structures can still raise an error from the DTA access or numeric functions.
Typical Call
[item, status] = labkit.dta.loadFile("pulse.DTA", "chrono");
assert(status.ok, status.message)
opts = struct( ...
"delay_s", 10e-6, ...
"cathLimit", -0.6, ...
"anodLimit", 0.8, ...
"area_cm2", 0.03, ...
"pulseMode", "Metadata first, then auto", ...
"usedMeasuredCurrent", true);
A = cic.analysisRun.computeCIC(item, opts);
assert(A.ok, A.message)
fprintf("Cathodic CIC: %.4g mC/cm^2\n", A.CICc_mCcm2)Related APIs
labkit.dta.detectPulses— Locate cathodic and anodic pulse windows in chrono data.vt_resistance.analysisRun.computeResistance— Estimate cathodic and anodic resistance from a VT pulse.
Source